丰满人妻一区二区三区无码AV,丰满少妇人妻HD高清大乳在线,丰满少妇人妻无码,丰满熟妇乱又伦在线无码视频

首頁 Home Product Equipments Microplate Reader

Microplate Reader

Analyte:

Samples:ELISA

Specification
HF4500
Product introduction
Summary

Device Overview

The new HF4500 microplate reader is developed based on extensive experiences in practical application. The instrument incorporates the latest technologies and strengths of similar domestic and international products. The instrument performs both single-wavelength and dual-wavelength assays, as well as qualitative and quantitative enzyme-linked immunoassays.

Equipment classification: Temporary pressured equipment; Category II; Pollution level: Class 2

Technical specifications

Light source: halogen, automatic calibration

Measurement channels: 4 vertical, optical channels; automatic centering

Wavelength range: 405 nm (optional), 450 nm, 492 nm, 630 nm; error of central wave:± 2 nm

Microplate type: 2X12 4X12 8X12

Measurement range: 0.000 to 4.000 OD

Resolution: 0.001 OD

Linear range: 0.000 to 3.000

Repeatability: 0.001

Accuracy: ±0.010

Display and input: 7-inch color LCD display; touch screen input?

Power input: wide voltage range of AC 220 V to 240 V, 50 to 60 Hz?

Product weight: Main unit has a net weight of 5.7 kg?

Dimensions: 390 mm (L) x 270 mm (W) x 155 mm (H)


Precautions
1) There should be a reliable power supply within the range of AC 220 V ± 10%, the frequency should be within 50 Hz ± 2%. Do not connect to 380 V voltages, and do not turn off the power during use. 2) Keep away from light, and store in a clean indoor environment away from strong electromagnetic fields. 3)?Maintain the required humidity (30% -80%), and maintain a relatively constant temperature in the room (20-25℃ is most suitable). 4)?Please use the standard 96-well ELISA plate for reading. 5)?Do not leave the micro-well plate and cuvette in the instrument slot for too long time after the test. Remove them from the instrument promptly after the test to prevent the solution from evaporating and corroding or damaging the internal optical path of the instrument. If a corrosive or volatile solution is tested, measure with a cover attached. 6)?Should the instrument behave abnormally or does not work properly, contact the manufacturer immediately. Do not disassemble or adjust the instrument. 7)??The instrument can be cleaned by wiping it with 2% alkaline or neutral glutaraldehyde solution or 0.5% chlorhexidine acetate-ethanol solution. 8)?In case of serious contamination, the instrument can be disinfected by the ethylene oxide disinfection method. 9)?Please follow the printer instructions to select acceptable printing paper.
More Products
Leaving a message
主站蜘蛛池模板: 兴隆县| 观塘区| 临邑县| 集贤县| 湖口县| 沁水县| 上饶市| 门头沟区| 霍林郭勒市| 玉门市| 阿拉善右旗| 和硕县| 武川县| 虹口区| 陇西县| 桃园县| 涟源市| 台中县| 芷江| 永昌县| 龙岩市| 得荣县| 安塞县| 德惠市| 韶山市| 蓝山县| 五大连池市| 革吉县| 谷城县| 临朐县| 庆云县| 滦平县| 沂源县| 福贡县| 珠海市| 湟中县| 浏阳市| 西藏| 昭觉县| 郯城县| 永丰县|